Wind energy slip rings with fault processing method for high and low temperature test Chamber

Edit: ByTune Electronics Co.,Ltd    Date: 2015-09-15
source:

Fault: temperature changes are not up to the test temperature values

Solution: check the electrical system, troubleshooting individually.

Breakdown II: temperature rise slowly or quickly

Solution: the temperature slowly check the air blowing system, look at the adjustment of air circulation flapper is open properly, and check the air circulation motor is normal. As temperatures rise more quickly you will need to set the parameters tuning of PID. If the temperatures rise directly, over temperature protection, then, controller failure, need to replace the control instrumentation.

Fault: low temperature not up to test indicators

Solution: observed temperature changes, temperature has a tendency to recover after a certain temperature, or temperature drop very slowly. If it is has a tendency to recover after a certain temperature the temperature was due to environmental equipment, equipment placed at ambient temperature, place of location (distance from the wall after the Cabinet) to meet the requirements (provided in the equipment operation instructions for use).

Fault four: damp heat test, actual and target humidity varies, many low numerical or actual humidity can reach 100%.

Solution: If the difference between the actual and target humidity is large, low a lot are mainly humidification system does not work, view the humidification system of the water supply system, is there a certain amount of water within the water supply system, humidification control water level control of boiler water level is normal, wet the water level in the boiler is normal.